Glass tube flow meter - TECFLUID 6000
+34 93 587 22 22 | WhatsApp : +34 628 512 590 grm@grmindustrial.es
Index » Productos » Instrumentation » Flow » Flow displayers » Glass tube flow meter

GLASS TUBE FLOW METER

FLANGES / THREAD

TECFLUID 6000

Range: 1/2 “- 3”

Accuracy: 1.6% (qG = 50%)

Maximum temperature: 80ºC.

Glass tube flow meter

GLASS TUBE FLOW DISPLAYER

FLANGES / THREAD

TECFLUID 6000

Range: 1/2 “- 3”

Accuracy: 1.6% (qG = 50%)

Maximum temperature: 80ºC.

Principle of operation

FLOW METER GLASS TUBE – TECFLUID 6000

The 6000 Tecfluid Series flow meters work according to the variable area principle, obtained by a float that moves inside a conical tube. borosilicate glass . The circulating flow moves to the float, up to a height inside the conical tube, where the forces are balanced: – E = fluid thrust – Pf = float weight – Al = free area of ​​passage (Al = Ac, area tube – Af, float area). Each float height or position corresponds to a value of the circulating flow.

Features

FLOW METER GLASS TUBE – TECFLUID 6000

Accuracy:

1,6% (qG=50%)

Materials:

  • Borosilicate glass tube. Parts in contact with the fluid in steel, AISI316L, PP, PVC and PTFE
  • Floats in AISI316L, aluminum, PTFE, PVDF-Pb, PVC-Pb, PTFE-Pb, titanium

Connection types:

Flange, thread, smooth in PVC for gluing and AISI 316, PP for welding and clamp

Sizes and flow rates:

  • From 1/2 “to 3”
  • From 2.5 l / h to 50 m3 / h H2O 20ºC

Maximum temperature:

80ºC

Maximum pressure:

Depending on the size of the meter, from 15 to 5 bar

Accessories:

  • Automation
  • Transmitter (ATEX certification and optional HART, PROFIBUS, FIELDBUS protocols.

Applications

FLOW METER GLASS TUBE – TECFLUID 6000

  • Water treatment
  • Heating and cooling flow control
  • Research laboratories
  • Gas burner control
  • Chemical industries and heat treatment furnaces
  • Test benches